VCT Vision ELITE Series Dokumentacja Strona 146

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 157
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 145
GE H
EALTHCARE
D
IRECTION
DOC0500644, R
EV
4 D
ISCOVERY
, L
IGHT
S
PEED AND
B
RIGHT
S
PEED
DICOM C
ONFORMANCE
S
TATEMENT
146 of 157
Attribute Name Tag Type Notes
BODY FILTER
MEDIUM FILTER
HEAD FILTER
Generator Power (0018,1170) 3 Always sent
Focal Spot (0018,1190) 3
Sent. Fixed value of 0.7 or
1.2.
Convolution Kernel (0018,1210) 3 Not sent for scouts.
Defined terms:
SMOOTH
SOFT
STANDARD
DETAIL
BONE
BONEPLUS
CHST
EDGE
SHARP
LUNG
Revolution Time (0018, 9305) 3 Sent for spiral scan only
Single Collimation Width (0018, 9306) 3 Sent for spiral scan only
Total Collimation Width (0018, 9307) 3 Sent for spiral scan only
Table Speed (0018, 9309) 3 Sent for spiral scan only
Table Feed per Rotation (0018, 9310) 3 Sent for spiral scan only
CT Pitch Factor (0018, 9311) 3 Sent for spiral scan only
It is possible for the operator of CT LightSpeed and BrightSpeed
systems to change the table height while scanning a series of
images. Therefore, implementations must use the Frame of
Reference UID (0020,0052) in conjunction with the Table Height
(0018,1130) to determine if two images are spatially related.
CAUTION