Przeglądanie stron 70
GE H
EALTHCARE
D
IRECTION
DOC0500644, R
EV
4 D
ISCOVERY
, L
IGHT
S
PEED AND
B
RIGHT
S
PEED
DICOM C
ONFORMANCE
S
TATEMENT
71 of 157
4.4 Communication Profiles
4.4.1 Supported Communication Stacks (parts 8,9)
DICOM Upper Layer (Part 8) is supported using TCP/IP.
4.4.1.1 TCP/IP Stack
The TCP/IP stack is inherited from the UNIX Operating System.
4.4.1.1.1 API
Not Applicable
4.4.1.1.2 Physical Media Support
Ethernet v2.0, IEEE 802.3.
4.5 Standard Extended / Specialized / Private SOPs
None.
4.6 Configuration
4.6.1 AE Title/Presentation Address Mapping
The Local AE title is "hostname_DCP", where “hostname” is the unique hostname within the
network.
The LightSpeed system allows the user to “add”, “Remove”, or “Update” the mapping of remote
DICOM Printer AE Titles to IP Addresses and Ports. These options can be selected from the
Service Desktop camera installation utility.
Note: If “hostname_DCP” is greater than 16 characters then the name of the local AE title is
limited to first 16 characters.
4.6.2 Configurable Parameters
The following fields are configurable for the DICOM Print SCU AE:
Note: All configurations should be performed by a GE Field Service Engineer.
Local IP Address
Local IP Netmask
Max PDU length
The following DICOM print parameters are configurable. The valid ranges are shown in
earlier sections of this conformance statement.
Medium Type (2000, 0030)
Film Destination (2000, 0040)
Przeglądanie stron 70