GE H
EALTHCARE
D
IRECTION
DOC0500644, R
EV
4 D
ISCOVERY
, L
IGHT
S
PEED AND
B
RIGHT
S
PEED
DICOM C
ONFORMANCE
S
TATEMENT
72 of 157
Magnification Type (2010, 0060)
Min Density (2010, 0120)
Max Density (2010, 0130)
Empty Image Density (2010, 0110)
Border Density (2010, 0100)
Configuration Information (2010, 0150)
Smoothing Type (2010, 0080)
The following network timers are supported and are configurable. When these timers expire
the association is terminated. The default values are:
Association Time out = 120 seconds
Session Time out = 1200 seconds (total time for open association)
N-SET Time out = 300 seconds
N-ACTION Time out = 300 seconds
N-CREATE Time out = 300 seconds
N-DELETE Time out = 300 seconds
N-GET Time out = 100 seconds
Time-outs, which are set for all hosts, are configurable in dprint.cfg:
*Association time-out - bi_assoc_tio
*Session timeout - bi_session_tio
*Create timeout - bi_ncreate_tio
Note: All configurations are performed by a GE Field Service Engineer.
4.7 Support of Extended Character Set
The Print SCU does not support any Extended Character Set.